Management of Chest Tube (ICD) with Air Leaks and Suction Settings

Table for Recap: Chest Tube (ICD) Management

Aspect

Action

Immediate Correction Needed

Initial Assessment

Ensure proper insertion and secure connections

Yes - Check for loose connections or improper setup

Water Seal Chamber

Verify appropriate water level

Yes - Fill to correct level if low

System Check (Tidaling)

Observe for fluctuations (tidaling)

Yes - Investigate if no movement

Identifying Air Leaks

Continuous bubbling indicates persistent air leak; intermittent may indicate minor leak

Yes - Assess and confirm source

External Cause Inspection

Inspect connections for looseness or disconnection

Yes - Tighten or replace loose connections

Tube Assessment

Ensure tube is not kinked or blocked

Yes - Fix or replace if blocked

Patient Assessment

Perform respiratory assessment and chest X-ray

Yes - Confirm tube position and assess complications

Initial Suction Setting

Start with -20 cmH2O

Yes - Adjust based on patient condition

Adjusting Suction

Incrementally increase suction by 5 cmH2O if needed

Yes - Reassess after each adjustment

Continuous Monitoring

Monitor respiratory rate, oxygen saturation, and overall status

Ongoing - Regular checks needed

Documentation

Record drainage amount, type, and presence of air leaks

Ongoing - Ensure accurate record-keeping

Chest Tube Position Re-eval

Use imaging to confirm correct placement

Yes - Correct if misplaced

Surgical Consultation

Consult thoracic surgeon if air leak persists

Yes - For persistent or severe leaks

Three-Bottle System

Use for managing severe air leaks

Yes - Ensure proper setup and usage

Patient Education

Educate on signs of complications (increased pain, difficulty breathing, etc.)

Ongoing - Provide clear instructions

System Maintenance

Regularly check for system function and integrity

Ongoing - Continuous maintenance required

Summary

  • Immediate Corrections: Focus on securing connections, checking water levels, ensuring proper tube positioning, and adjusting suction settings.

  • Ongoing Actions: Regular patient assessment, monitoring, documentation, and maintenance of the chest tube system.

  • Consultations: Engage thoracic surgeons for persistent or severe air leaks.

This table helps to quickly identify what aspects of chest tube management can be regularly maintained and what requires urgent action.


 

Chest tubes, also known as intercostal drainage (ICD), are essential for managing conditions where air or fluid accumulates in the pleural space. The presence of air bubbles in the water seal chamber indicates an air leak, which can originate either from the chest tube system or the patient's lung. Here is a comprehensive guide for managing such scenarios:

Initial Assessment and Setup

  • Ensure Proper Insertion and Security:

    • Confirm that the chest tube is properly inserted and secured.

    • Verify that the chest tube connections are airtight. Loose connections can cause air leaks.

    • Check that the water seal chamber is filled to the appropriate level as per the manufacturer’s guidelines.

  • System Check (Fluctuation - Tidaling):

    • Observe the water level in the water seal chamber for fluctuations (tidaling). This movement indicates that the system is functioning and the pleural space is responding to respiratory cycles.

      • Normal fluctuation: Indicates proper system function.

      • No fluctuation: May indicate a blocked tube, re-expansion of the lung, or disconnection. Investigate further if no movement is observed.

Identifying the Source of Air Leaks

  • Bubbling in the Water Seal Chamber:

    • Continuous Bubbling: Indicates a persistent air leak, often due to a bronchopleural fistula or lung parenchymal injury.

    • Intermittent Bubbling: May be due to the patient’s respiratory cycle, suggesting a minor leak.

Troubleshooting and Management

  • Check for External Causes:

    • Inspect all connections for any signs of looseness or disconnection.

    • Ensure the tube is not kinked or blocked.

    • Replace or tighten any loose connections.

  • Assess the Patient:

    • Perform a thorough respiratory assessment, including checking breath sounds and respiratory rate.

    • Use a chest X-ray to confirm the tube’s position and detect any residual pneumothorax or other complications.

Suction Management

  • Suction Settings:

    • Typically, the suction pressure is set between -10 cmH2O and -20 cmH2O.

    • Increase suction: If the air leak persists and the patient’s condition requires it, suction may be increased to up to -40 cmH2O, but this should be done cautiously and based on clinical need.

  • Detailed Steps for Suction Application:

    • Initial Suction Application:

      • Start with a suction setting of -20 cmH2O.

      • Monitor for changes in bubbling and the patient’s respiratory status.

    • Adjusting Suction:

      • If continuous bubbling persists, and there are no signs of tube dislodgment or blockage, consider increasing suction incrementally by 5 cmH2O, reassessing after each adjustment.

Patient Monitoring

  • Continuous Monitoring:

    • Monitor the patient's respiratory rate, oxygen saturation, and overall respiratory status.

    • Regularly check the drainage system to ensure it is functioning correctly and there is no blockage or disconnection.

  • Documentation:

    • Record the amount and type of drainage, presence of air leaks, and any changes in the patient's condition.

Interventions for Persistent Air Leaks

  • Re-evaluate Chest Tube Position:

    • Use imaging to confirm the correct placement of the chest tube.

  • Surgical Consultation:

    • If the air leak persists despite increased suction, consult with a thoracic surgeon. Surgical intervention may be necessary for cases involving significant lung damage or bronchopleural fistula.

Additional Considerations

  • Using a Three-Bottle System:

    • This system can manage severe air leaks by providing separate compartments for suction control, water seal, and drainage collection.

      • First Bottle (Drainage Collection): Collects fluid directly from the pleural space.

      • Second Bottle (Water Seal): Acts as a one-way valve to prevent air from entering the pleural cavity.

      • Third Bottle (Suction Control): Regulates the amount of suction applied to the system.

Conclusion

Managing a chest tube with an air leak involves a systematic approach to ensure patient safety and effective resolution of the issue. Proper initial setup and continuous monitoring are critical for the effective management of a chest tube. Adjusting suction based on clinical need, ensuring all connections are secure, and performing regular assessments can help manage and resolve air leaks. In cases of persistent or severe air leaks, surgical consultation may be required. Regular assessment, proper setup, and timely interventions are essential for optimal patient outcomes.